Status:

UNKNOWN

Safety of Cardiac Pacemakers in 3 Tesla MRI

Lead Sponsor:

University Hospital, Bonn

Conditions:

Heart Diseases

Eligibility:

All Genders

Brief Summary

To evaluate short-term and long-term effects by MR imaging at 3T on the technical and functional status of cardiac pacemakers.

Detailed Description

Recently, a shift could be observed from cardiac pacemakers being an absolute contraindication to being a relative contraindication for MR imaging at 1.5 Tesla, depending on a patient-per-patient risk...

Eligibility Criteria

Inclusion

  • Urgent need for an MRI examination
  • Stable pacemaker physical parameters
  • Battery voltage \> 2.7 V
  • Battery impedance \< 2000 Ohm
  • Battery estimated remaining lifetime \> 6 months
  • Lead impedances 200-2000 Ohm
  • Stable pacing parameters
  • Pacing capture threshold \<2.5V at a pulse duration of 0.4ms
  • Sensing \> 5mV
  • Minimum 3 months since pacemaker and lead implantation

Exclusion

  • History of ventricular tachycardia (VT) or ventricular fibrillation
  • Unstable angina
  • Myocardial infarction within the previous 3 months
  • Cardiothoracic surgery within the previous 3 months
  • Presence of an abdominal pacemaker with lead length \> 70 cm
  • Presence of MRI incompatible bioimplants or other MRI incompatible materials

Key Trial Info

Start Date :

Trial Type :

OBSERVATIONAL

End Date :

Estimated Enrollment :

75 Patients enrolled

Trial Details

Trial ID

NCT00356330

Last Update

April 11 2007

Active Locations (1)

Enter a location and click search to find clinical trials sorted by distance.

Page 1 of 1 (1 locations)

1

University of Bonn, Department of Radiology

Bonn, North Rhine-Westphalia, Germany, 53127